ChemNet
 
Previous article Next article Contents  

Andrey V. Alekseev, Pavel V. Yakimovich

Estimation of uncertainty budgets in analysis of nickel by ICP-MS and high-resolution mass spectrometry with a glow discharge

Abstract

Abstract.The paper estimates uncertainty budgets in determining the content of impurities of 15 elements (P, Mn, Fe, Cu, Zn, As, Se, Ag, Cd, Sn, Sb, Te, Tl, Pb, Bi) in nickel samples by mass spectrometry inductively coupled plasma (ICP-MS) and high-resolution glow discharge mass spectrometry (GDMS). Validated by analysis of a certified standard nickel sample (assessed using Student’s simple t-test). The main sources of uncertainty in the results of analysis by two methods, including sample preparation and calibration graph, are presented. Comparison of the total and expanded uncertainties for the two methods is carried out. The analysis by ICP-MS and GDMS methods of industrial melting of a nickel alloy of the ZhS6U type was carried out and the results obtained were compared with the modified Student’s test.
Key words: inductively coupled plasma mass spectrometry, glow discharge mass spectrometry, nickel alloys, impurities determination, nickel analysis, uncertainty budget
Moscow University Chemistry Bulletin.
2022, Vol. 63, No. 4, P. 283
   

Copyright (C) Chemistry Dept., Moscow State University, 2002
   Overview
   Editorial board
   Tables of Contents
   Subscription

The site is supported by Russian Foundation for Basic Research
  The using of published on this page materials is not allowed without special permission
Copyright (C) Chemisty Department of Moscow State University
Web-Editor: B.I.Pokrovskii
Web-design: Copyright (C) MIG and VVM
webmaster@www.chem.msu.su